10-30 Section 10-8 - Electrical Safety Tests
10-8-13 Total Patient leakage current : Mains to Type CF probe (Only for Europe)
Reference the procedure in the IEC 60601-1 Ed 3. Measure leakage current flow from mains to all ECG
leads connected together.
Table 10-31 Typical Data Sheet for Total Patient Leakage Current - Mains to Type CF probes
DANGER
!! DANGER:
Electric Shock Hazard.
Line voltage is applied to the ECG leads during this test. To avoid possible electric shock
hazard, the system being tested must not be touched by patients, users or anyone during
testing.
CAUTION
!! CAUTION:
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Tester Neutral Tester Polarity Switch Ground Switch Measured Leakage Current
Closed Normal Closed
Closed Reversed Closed

Table of Contents

Other manuals for GE Voluson P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 and is the answer not in the manual?

GE Voluson P6 Specifications

General IconGeneral
BrandGE
ModelVoluson P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals