GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-123
Section 8-39
Replacement of the REAR HANDLE
8-39-1 Manpower
One person, 5 minutes
8-39-2 Tools
Philips screwdriver, Allen Wrench
8-39-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-39-4 Removal Procedure
1.) Remove the ASSY BEZEL FRONT PRINTER PLASTIC BT16. Refer to the figure below.
Figure 8-179 Removing the ASSY BEZEL FRONT PRINTER PLASTIC BT16
ASSY BEZEL FRONT PRINTER
PLASTIC BT16