GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-24 Section 8-7 - Replacement or Activation of Options
NOTE: Option a) : Can be used with 3D/4D Activation
8-7-1 Operation for activating Options
1.) Press the F2 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION
and then click the OPTIONS tab (see:
Figure 8-2 on page 8-7.)
8-7-1-1 Operation for installing a “Demo Key” or a “Permanent Key”:
1.) Position the cursor inside the input field desired and press the right/left trackball
key.
2.) If one exists, clear/edit the current key code.
3.) Enter the encrypted serial code with the keyboard and click on SUBMIT
. (Code will be checked.)
4.) Click the SAVE&EXIT
button.
NOTE: After activating a key code, restart (turn off and on) the Voluson™ P8/Voluson™P6 system.
Elastography N/A N/A Standard N/A N/A
B-Flow Standard Standard Option N/A N/A
Advanced SRI Standard Standard Standard N/A N/A
CFM/M-CFM Standard Standard Standard Standard Option
DICOM Standard Standard Standard Option Option
Activate 4th Probe Port Standard Standard Option N/A N/A
Scan Assistant Standard Standard Option N/A N/A
3D Printer Export (Full Mesh) Standard Standard Option Option Option
E8C-RS Activation (for Japan only) N/A N/A N/A N/A Option
Veterinary Option Option Option Option Option
Table 8-4 Software Option
S/W Options S10 Expert S10 S8 P8 P6