EasyManua.ls Logo

GE Voluson P6 User Manual

GE Voluson P6
620 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-3
Section 5-2 General Information (cont?)
Among other significant features of the Voluson™ P8/Voluson™P6 are the following:
- Integrated FrontEnd (uses advanced FGPA technologies)
- Open connectivity using USB ports
- Bluetooth - wireless connectivity
- High performance 17” LCD monitor
- Low profile, backlit Keyboard
Voluson™ P8/Voluson™P6 has a digital beam forming system (incorporated in the FrontEnd) which
can handle up to 192 element probes by use of multiplexing.
Signal flow from the Probe Connector Panel, to the FrontEnd (FE) Electronics, to the BackEnd
Processor, and finally is displayed on the LCD monitor and peripherals.
Voluson™ P8/Voluson™P6 internal electronics are divided into three:
- Front End (FE) Processor
- Back End Processor
- Power Supply Unit
Interconnecting signals from FrontEnd, BackEnd, keyboard, monitor, and power distribution sub-
systems are routed internally.
Major System Components:
FrontEnd processor: Section 5-3 on page 5-25
BackEnd processor: 5-4-2 on page 5-31
Top Console User interface (System I/O with hard keys):
Section 5-6 on page 5-34
•Monitor: Section 5-7 on page 5-40
External I/O: Section 5-9 on page 5-41
Peripherals: Section 5-10 on page 5-42
System mechanical chassis: trolley to keep all major components Figure 5-1 on page 5-2

Table of Contents

Other manuals for GE Voluson P6

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 and is the answer not in the manual?

GE Voluson P6 Specifications

General IconGeneral
Transducer Ports3
3D/4D ImagingYes
Power Requirements100-240 VAC, 50/60 Hz
Storage Temperature-20°C to 60°C
ApplicationObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
Probe TypesConvex, Linear, Endocavity, Phased Array
Operating Temperature10°C to 40°C
Humidity Range30% to 80%
Transducer Optionsendocavity probes