GPT-9000/9000A Series User Manual
128
MEASure<x>
Query
Description
Returns the test parameters & results of the tester
in either MANU or AUTO mode.
MANU mode: Returns the test parameters &
results of a MANU test.
AUTO mode: Returns the test parameters & results
of the selected step (1-16) of the AUTO test.
Return parameters: function, judgment/status, test
voltage, test current/resistance, test time (time of
completed test) or ramp time (elapsed time of test
that has not been completed.
Query Syntax
MEASure<x>?
Parameter
(MANU mode)
No parameter needed for MANU
mode.
Parameter
(AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter
<string>
Returns the test status of the test
in the following format:
function, judgment or status, test
voltage, test current or resistance,
test time or ramp time
Function
ACW, DCW, IR, GB
Judgment
/Status
PASS, FAIL
VIEW
Test voltage
voltage+unit
Test current
/Test resistance
current+unit
resistance+unit
Test time
/Ramp time
T=time+S
R=time+S
Example
(in MANU mode)
MEAS?
>ACW,FAIL,0.024kV,0.013 mA ,R=000.1S
Returns the test result of the current manual test.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals