GPT-9000/9000A Series User Manual
136
MANU:ACW:FREQuency
Set
Query
Description
Sets or returns the ACW test frequency in Hz. The
test must first be in ACW mode before this
command can be used.
Syntax
Query Syntax
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/
Return parameter
50
50 Hz
60
60 Hz
Example
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
MANU:ACW:REF
Set
Query
Description
Sets or returns the ACW reference value in mA.
The test must first be in ACW mode before this
command can be used.
The ACW reference value must be less than the HI
SET value.
The ACW reference value must use the same range
as the HI SET value.
Syntax
Query Syntax
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX/99XXA)
Example
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals