GPT-9000/9000A Series User Manual
136
Sets or returns the ACW test frequency in Hz. The
test must first be in ACW mode before this
command can be used.
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/
Return parameter
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
Sets or returns the ACW reference value in mA.
The test must first be in ACW mode before this
command can be used.
The ACW reference value must be less than the HI
SET value.
The ACW reference value must use the same range
as the HI SET value.
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/
Return parameter
0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX/99XXA)
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.