OPERATION
55
Setting PASS HOLD
The PASS HOLD settings only apply to the
selected test in an AUTO test. When the PASS
HOLD setting is set to ON, a PASS judgment is
held until the START button is pressed.
The PASS HOLD setting only applies to AUTO
tests. This setting is ignored when running a
single MANU test.
1. Press the UTILITY key on the front
panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.
L I T Y IM A UN U T 2M A N U = * * * - 0 0
A R C M D E : OO FF
G R O U N D MO D E : NO
P A S S H L D :O O FF
F A I L M D E :O
PS T O
M A X H O L D : O F F
The MANU UTILITY settings only apply to the
selected MANU test.
2. Use the UP/DOWN arrow keys to
move to the PASS HOLD setting.
3. Use the scroll wheel to set PASS
HOLD.
4. Press the EDIT/SAVE key to save
and exit the MANU Utility menu.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com