GPT-9000/9000A Series User Manual
132
MANU:INITial
Loads the initial settings for the selected MANU
number.
Sets or returns the Ramp Time for the test in
seconds.
Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is ≥ 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function
only.
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/
Return parameter
MANU:RTIM 0.5
Sets the ramp time to half a second.
Sets or returns the mode (ACW, DCW, IR, GB) of
the selected manual test.
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?