Sets or returns the IR reference value in MΩ (GPT-
98XX) or GΩ. The test must first be in IR mode
before this command can be used.
The reference value must be lower than the HI SET
value.
MANU:IR:REF <NR1>
MANU:IR:REF?
Parameter/
Return parameter
GPT-98XX only:
0000 ~ 9999 (unit = MΩ)
GPT-99XX/GPT-99XXA only:
Format A: 0 ~ 50.00 (unit = GΩ)
Format B: 0G ~ 50.00G
Format C: 0M ~ 50000M
MANU:IR:REF 900
Sets the IR reference to 900 MΩ.
Example
(GPT-99XX/
99XXA)
MANU:IR:REF 0.900
Sets the IR reference to 900 MΩ.
Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/
Return parameter
3.00~33.00 (GPT-98XX)
3.00~33.00 (GPT-99XX)
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.