GPT-9000/9000A Series User Manual
134
MANU:ACW:CHISet
Set
Query
Description
Sets or returns the ACW HI SET current value in
milliamps. The test must first be in ACW mode
before this command can be used.
Syntax
Query Syntax
MANU:ACW:CHISet <NR2>
MANU:ACW:CHISet?
Parameter/
Return parameter
<NR2>
0.001 ~ 042.0 (GPT-98XX)
0.001 ~ 110.0 (GPT-99XX/99XXA)
Example
MANU:ACW:CHIS 10.0
Sets the ACW HI SET current to 10 mA.
MANU:ACW:CLOSet
Set
Query
Description
Sets or returns the ACW LO SET current value in
milliamps. The LO SET value must be less than the
HI SET value. The test must first be in ACW mode
before this command can be used.
The LO SET range must use the HI SET range. If all
the digits in the LO SET range are outside the HI
SET range, an error will be produced. All digits
outside the HI SET range are ignored and will not
be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals