MANU:GB:RHISet
Set
Query
Description
Sets or returns the GB HI SET resistance value in
m. The test must first be in GB mode before this
command can be used.
Syntax
Query Syntax
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/
Return parameter
<NR2>
000.1 ~ 650.0
Example
MANU:GB:RHIS 100.0
Sets the HI SET value to 100m.
Note
If the (GB current x HI SET resistance) > 5.4V, then an
error will be generated (“GBV > 5.4V”).
MANU:GB:RLOSet
Set
Query
Description
Sets or returns the GB LO SET resistance value in
m. The LO SET value must be less than the HI
SET value. The test must first be in GB mode
before this command can be used.
Syntax
Query Syntax
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50m.
MANU:GB:TTIMe
Set
Query
Description
Sets or returns the GB test time in seconds. The test
must first be in GB mode before this command can
be used.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals