Sets or returns the GB HI SET resistance value in
mΩ. The test must first be in GB mode before this
command can be used.
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/
Return parameter
MANU:GB:RHIS 100.0
Sets the HI SET value to 100mΩ.
If the (GB current x HI SET resistance) > 5.4V, then an
error will be generated (“GBV > 5.4V”).
Sets or returns the GB LO SET resistance value in
mΩ. The LO SET value must be less than the HI
SET value. The test must first be in GB mode
before this command can be used.
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/
Return parameter
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50mΩ.
Sets or returns the GB test time in seconds. The test
must first be in GB mode before this command can
be used.