5.5.2 Initial Startup Window (During Initialization)
Listed below are principally messages that are displayed when an application is being started
(when the initial window is being displayed).
<<Error Message List>>
Code Error Message Countermeasure
1101 The data in Adjust File are
wrong.
Cannot start up the application.
Click the [OK] button to close the program.
Adjustment file “TM_Adjust.ini” contains no data,
or some of the data are missing. Copy the
adjustment file.
For adjustment file copying, refer to 2.5.3 in the
instruction manual.
1111 No data found in Initial File.
Do you default to start up PC?
Initial file contains no data or some of the data
are missing. To start this application in default
setting, click the [OK] button, or the [Cancel]
button to terminate this application.
After the application starts, set observation
screen size (referring to 2.5.7 in instruction
manual).
1112 The data in Initial File are
wrong.
Do you default to start up PC?
Initial file contains abnormal data or some of the
data are missing. To start this application in
default setting, click the [OK] button, and the
[Cancel] button to terminate this application.
After start of the application, set observation
window size (referring to 2.5.7 in instruction
manual).
1113 Cannot start up the application.
Failed to access Initial File.
If this application could not be started in default
setting, this message is displayed. When the
account type into computer is Limited user,
switch to the other account as Computer
Administrator again. Set the observation window
size referring to 2.5.7 in the instruction manual.
Then it is possible to start this application for
Limited user.
1115 Set up the observation screen
size when the application has
started.
Click the [OK] button to close the program.
Set the observation window size. For details,
refer to 2.5.7 in the instruction manual.