2-2
Installation
Operating Environment
Operating Environment
This instrument is designed for Indoor use only.
DO NOT operate the product in an explosive atmosphere or in the
presence of flammable gasses or fumes.
This instrument may be operated in environments within the following
limits:
Temperature: 0
o
C to +45
o
C
+5
o
C to +40
o
C for Jitter operation.
+5
o
C to +35
o
C with lid printer
Altitude up to 3050m (10,000ft)
Humidity up to 95% relative humidity to 40
o
C, but it should be
protected from temperature extremes which may cause
condensation.
To ensure adequate cooling do not obstruct air vents in the instrument
cabinet. Do not for example operate the instrument if it is standing on its
rear feet, as air vents may be obstructed by floor covering.
CAUTION This instrument is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 644 respectively.
CAUTION VENTILATION REQUIREMENTS: When installing the instrument in a
cabinet, the convection into and out of the instrument must not be
restricted. If the total power dissipated in the cabinet is greater than 800
watts, then forced convection must be used.