2-4
Installation
Preparation for Use
WARNING BEFORE REMOVING THE FUSE, THE AC LINE POWER CORD
SHOULD BE DISCONNECTED FROM THE POWER SOURCE
AND THE OTHER END DISCONNECTED FROM THE
INSTRUMENT.
WARNING ONLY USE A FUSE OF THE CORRECT RATING AS LISTED IN
“Fuses” on page 2-3. DO NOT use repaired fuses or short-circuited
fuseholders: For continued protection against fire, replace the line fuse(s)
only with fuse(s) of the same voltage and current rating and type.
The fuse is removed by inserting a suitable flat bladed tool into the slot
in the fuse cap and turning counter-clockwise. The cap and the fuse can
then be removed and the fuse changed for another of the correct rating.
The fuse rating and HP part number are listed in “Fuses” on page 2-3.
Connecting to the Power Supply
WARNING This is a Safety Class I instrument (provided with a protective earthing ground,
incorporated in the powercord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the instrument is likely to make the
instrument dangerous. Intentional interruption is prohibited.
CAUTION Before switching on this instrument, make sure that the line supply voltage is in the
specified ranges. Range selection is automatic.