2100-900-01 Rev. D / September 2011 Return to Section Topics B-9
Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual Appendix B: Remote Interface Reference
CONFigure:DIODe
Preset and configure for a diode measurement. This command does not initiate the measurement.
The range and resolution are fixed at 1VDC with 1mA current source and 5 1/2 digits respectively.
CONFigure:TEMPerature
Preset and configure for an RTD measurement. This command does not initiate the measurement.
The resolution is fixed at 6 1/2 digits.
CONFigure?
Query a recent condition of a function and return a quoted string.
The measurement configuration command
[SENSe:]FUNCtion "<function>"
Select a measurement function and enclose it in quotes in the command string (FUNC
"VOLT:DC"). Use one of the following strings:
VOLTage:DC
VOLTage:AC
VOLTage:DC:RATio
CURRent:DC
CURRent:AC
RESistance (for 2-wire ohms)
FRESistance (for 4-wire ohms)
FREQuency
PERiod
CONTinuity
DIODe
TEMPerature
[SENSe:]FUNCtion?
Query the measurement function and return a quoted string.
[SENSe:]<function>:RANGe {<range>|MINimum|MAXimum}
Select a range for the selected function. For frequency and period measurements, range applies
to the signal's input voltage, not its frequency (use FREQuency:VOLTage or PERiod:VOLTage).
MIN selects the lowest range for the selected function. MAX selects the highest range.