B-14 Return to Section Topics 2100-900-01 Rev. D / September 2011
Appendix B: Remote Interface Reference Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual
[SENSe:]ZERO:AUTO?
Query the AUTO ZERO mode. Returns "1" (ON) or "0" (OFF or ONCE).
[SENSe:]GAIN:AUTO {OFF|ONCE|ON}
Disable or enable the AUTO GAIN mode. The OFF and ONCE have a similar effect. OFF mode
does not issue a new offset measurement until the multimeter goes to the "WAIT-FOR-TRIGGER"
state. Parameter ONCE issues an immediate offset measurement.
[SENSe:]GAIN:AUTO?
Query the AUTO GAIN mode. Returns "1" (ON) or "0" (OFF or ONCE).
INPut:IMPedance:AUTO{OFF|ON}
Disable or enable the automatic input resistance mode for DC voltage measurements. With AUTO
ON, the input resistance is set to >10G for the 100mV, 1V, and 10V ranges. With AUTO OFF, the
input resistance is fixed at 10M for all ranges.
INPut:IMPedance:AUTO?
Query the input resistance mode. Returns "1" (ON) or "0" (OFF).
ROUTe:TERMinals?
Query the multimeter to determine if the front or rear input terminals are selected. Returns "FRON"
or "REAR."
The math operation command
CALCulate:FUNCtion {PERCent|AVERage|NULL|LIMit|MXB|DB|DBM}
Select the math function (only one function can be enabled at a time. The default function is
PERCent).
CALCulate:FUNCtion?
Query the present math function. Returns PERC, AVER, NULL, LIM, MXB, DB or DBM.
CALCulate:STATe {OFF|ON}
Disable or enable the selected math function.