4-6 Return to Section Topics 2100-900-01 Rev. D / September 2011
Section 4: Front Panel Operations Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual
To set the number of readings from the front panel:
1. Press the CONFIG key, and then the FILTER key.
2. Use the PREV and NEXT keys to select the READINGS submenu, then press ENTER.
3. Use the up and down arrow keys to enter the number of readings (from 2 to 100) to be
taken in order to get an average value. Press the ENTER key when you are prompted to
confirm your settings.
To set the mode from the front panel:
1. Press the CONFIG key, and then the FILTER key.
2. Use the PREV and NEXT keys to select the MODE and press the ENTER key.
3. Use the PREV and NEXT keys to select MOVING AVG or REPEAT AVG, then press the
ENTER key to confirm your settings.
4. Press the FILTER key to enable or disable this function.
To set the digital filter from the remote interface:
Use the following commands to set your digital filter:
SENSe:AVERage:TCONtrol {MOVing|REPeat}
SENSe:AVERage:TCONtrol?
SENSe:AVERage:COUNt {<value>|MINimum|MAXimum}
SENSe:AVERage:COUNt? [MINimum|MAXimum]
SENSe:AVERage:STATe {OFF|ON}
SENSe:AVERage:STATe?
Resolution setting (digits)
Definition
Resolution is the number of digits a multimeter can measure. Using the Model 2100 multimeter,
you can select the resolution for a specific measurement. The choices for the resolution setting
are: fast 4 1/2, slow 4 1/2, fast 5 1/2, slow 5 1/2, fast 6 1/2, and slow 6 1/2. For higher
measurement accuracy, select 6 1/2-digit resolution; for faster measurements, select 4 1/2-digit
resolution.
The resolution setting applies to all math operations under the selected measurement function.
The selected value is stored in a volatile memory and the setting is only valid for the current
measurement function. You can select different resolutions for different measurement functions.
Default
The default value for resolution is “slow 5 1/2 digit” and the default value will be restored when the
meter has been turned off or after a remote interface reset. The user-selected value for the
resolution setting is stored in a volatile memory and the setting is only valid for the current
measurement function (refer to
Table 2-2).
NOTE The resolution setting for AC measurements is actually fixed at 6 1/2 digits. When you
select a lower resolution, the meter will mask out the extra digits. Your selection of
resolution for AC function will not affect the actual speed or accuracy.
How to set the resolution
The resolution can be set either through the front panel or a remote interface.