2100-900-01 Rev. D / September 2011 Return to Section Topics B-19
Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual Appendix B: Remote Interface Reference
READ?
Change the state of the triggering system from the "idle" state to the "wait-for-trigger" state. The
meter will start to make measurements when a required triggering condition is met after the
READ? command is received. Measurement readings are sent to the output buffer immediately.
DISPlay {OFF|ON}
Turn off or on the display.
DISPlay?
Query the display setting. Returns "0" (OFF) or "1" (ON).
DISPlay:TEXT <quoted string>
Show a message on the front panel display. The allowed message can be up to 16 characters in
the lower row display; any additional characters will be truncated.
DISPlay:TEXT?
Query the message sent to the front panel display.
DISPlay:TEXT:CLEar
Clear the message shown on the front panel display.
SYSTem:BEEPer
Issue a single beeping tone.
SYSTem:BE80
Query the beeper mode. Returns "0" (OFF) or "1" (ON).
SYSTem:ERRor?
Query the multimeter's error queue. Up to 20 errors can be stored in the queue. Errors are
retrieved in first-in-first-out (FIFO) order. Each error string may contain up to 80 characters.
SYSTem:VERSion?
Query the present SCPI version.