8-4 Measurement
îš„ Abdomen measurements - Used for measurements of abdominal organs (liver, gall
bladder, pancreas and kidney, etc.) and large abdominal vessels.
îš„ OB measurements - Used for measurements of fetal growth indexes (including EFW) as
well as GA and EDD calculations. The fetus can be evaluated through growth graph
analysis and fetal biophysical profile.
 Cardiac measurements – Used for left ventricle function measurements and
measurements of main artery and vein parameters, etc.
îš„ Gynecology measurements - Used for the uterus, ovary and follicles, etc.
 Small Part measurements – Used for small parts such as thyroid.
îš„ Urology measurements - Used for prostate, seminal vesicle, renal, adrenal, micturated
and testicle volume.
 Orthopedics measurements – Used for hip joint measurement.
 Vascular measurements – Used for carotid, cerebral, upper and lower extremities vessels,
etc.
 EM measurements – Used for EM related full functional measurement.
8.4 Measurement Accuracy
Table 1 Error of 2D Images
Parameter Value Range Error
Distance Full screen Within ±2%
Area Full screen Within ±6%
Circ Full screen Within ±7%
Angle Full screen Within ±3%.
Volume Full screen Within ±7 %
Table 2 iScape Measurements
Parameter Probe Type Error
Distance
Full screen Within ±5 % (L14-6NE, L12-3E, P4-2, D7-
2E, 3C5A, 7L4A probe)
Full screen Within ±10 % (V11-3, CB10-4E probe)
Table 3 Time/Motion Measurements
Parameter Value Range Error
Distance Full screen Within ±3 %
Time Timeline Display Within ±2%
Heart rate Timeline Display Within ±3%
Velocity (PW
mode)
10-200cm/s: (for
L14-6NE, L12-3E,
CB10-4E, 7L4A)
10-300cm/s: (for
P4-2, D7-2E, V11-3,
3C5A)
When angle≤60º, ≤5%.