Safety Precautions 1-9
10. The damage of the probe may be caused by the contact of
improper gel or cleaner:
z DO NOT soak or saturate probes in the strong polar solution
of ethanol, chloride of lime, ammonium chloride, acetone or
formaldehyde.
z DO NOT contact the probe with solutions or ultrasound gels
containing oily medium such as mineral oil or lanoline.
NOTE: Read the following precautions to prevent the probe from malfunction:
z Before connecting or disconnecting the probe, freeze or turn off the
diagnostic ultrasound system.
z Clean and disinfect the probe before and after each examination.
z After the examination, wipe off the ultrasound gel thoroughly. Otherwise,
the ultrasound
gel may solidify and the image quality would be degraded.
Ambient conditions:
1. To prevent the probe from being damaged, do not use it where it will be
exposed to:
z Direct sunlight or X-rays
z Sudden changes in temperature
z Dust
z Excessive vibration
z Heat generators
2. Use probes under the following ambient conditions :
z ambient temperature: 0°C to 40°C
z relative humidity: 30% to 85% (no condensation)
z atmospheric pressure: 700 hPa to 1060 hPa
3. Use D7-2E, D6-2NE probes under the following ambient conditions :
z ambient temperature: 10°C to 40°C
z relative humidity: 30% to 85% (no condensation)
z atmospheric pressure: 700 hPa to 1060 hPa
Repeated disinfection will eventually damage the probe, please check the probe
performance periodically.
NOTE: The following definition of the WEEE label applies to EU
member states only: The use of this symbol indicates that this
system should not be treated as household waste. By ensuring
that this system is disposed of correctly, you will help prevent
bringing potential negative consequences to the environment
and human health. For more detailed information with regard to
returning and recycling this system, please consult the
distributor from whom you purchased the system.
For system products, this label may be attached to the main unit only.
1.5 Latex Alert
When choosing a probe sheath, it is recommended that you directly contact CIVCO for
obtaining probe sheath, pricing information, samples and local distribution information. For
CIVCO information, please contact the following:
CIVCO Medical Instruments
Tel: 1-800-445-6741

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-45 and is the answer not in the manual?

Mindray DC-45 Specifications

General IconGeneral
BrandMindray
ModelDC-45
CategoryMedical Equipment
LanguageEnglish

Related product manuals