EasyManuals Logo
Home>Mindray>Medical Equipment>DC-8 EXP

Mindray DC-8 EXP User Manual

Mindray DC-8 EXP
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
Urology 9-5
NOTE: Need to be measured in left and right side respectively.
1. Select [Testis Vol] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure Testis L,
Testis H and Testis W. The Testis Vol is calculated automatically.
Pre-BL Vol
Function: measures the Pre-BL L, Pre-BL H and Pre-BL W, calculates the Pre-BL Vol.
1. Select [Pre-BL Vol] in the measurement menu.
2. Use the Distance method in 2D General Measurements to measure Pre-BL L, Pre-BL H and
Pre-BL W. The Pre-BL Vol is calculated automatically. The Mictur.Vol is displayed in the report
if the Post-BL Vol is measured.
Post-BL Vol
Function: measures the Post-BL L, Post-BL H and Post-BL W, calculates the Post-BL Vol.
1. Select [Post-BL Vol] in the measurement menu.
2. Use the Distance method in 2D General Measurements to measure Post-BL L, Post-BL H and
Post-BL W. The Post-BL Vol is calculated automatically. The Mictur.Vol is displayed in the
report if the Pre-BL Vol is measured.
Mictur.Vol
Function: measures the Pre-BL Vol and Post-BL Vol, calculates the Mictur.Vol.
1. Select [Mictur.Vol] in the measurement menu.
2. Use the Distance method in 2D General Measurements to measure Pre-BL L, Pre-BL H and
Pre-BL W. The Pre-BL Vol is calculated automatically.
3. Use the Distance method in 2D General Measurements to measure Post-BL L, Post-BL H and
Post-BL W. The Post-BL Vol and Mictur.Vol are calculated automatically.
9.4.3 Study Tool Operations
Kidney
Function: measures Renal L, Renal H and Renal W, calculates Renal Vol.
NOTE: Need to be measured in left and right side respectively.
1. Select the [Kidney] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure Renal L,
Renal H and Renal W. The Renal Vol is calculated automatically.
3. Use the method of Distance measurement of 2D General Measurements to measure Cortex.
Adrenal
Function: measures Adrenal L, Adrenal H and Adrenal W.
NOTE: Need to be measured in left and right side respectively.
1. Select the [Adrenal] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure Adrenal
L, Adrenal H and Adrenal W.
Prostate
Function: measures Prostate L, Prostate H and Prostate W, calculates Prostate Vol and PPSA. If
[Serum PSA] in [Patient Info] -> [URO] has been input, PSAD (Prostate Special Antigen Density)
will also be calculated.
PPSA (ng/ml) = PPSA Coefficient (ng/ml
2
) × Prostate Vol (ml)

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish