Chapter 4 Reprocessing Operations
107
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6. Close the lid to drain the cleaning fluid and finish leak test. At this time, the
buzzer beeps and the main control panel displays [- -] to indicate the end of
testing.
Figure 4.30
Once the leak test air tube has been connected to a
nonreprocessed endoscope, the outer surfaces of the tube
should be disinfected. In this case, the leak test air tube
should remain connected to the endoscope for the entire
reprocessing cycle.
When 10 minutes have passed since the leak test started,
the water will drain and the test will finish automatically. At
this time, the main control panel displays the error code
[E92].