226
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Do not add fresh disinfectant solution to the used disinfectant
solution when the disinfectant solution in the device is no
longer effective or the volume of the disinfectant solution in
the device drops below its usable level. Drain the disinfectant
solution completely and replace with fresh disinfectant
solution.
Draining the disinfectant solution
Before draining the disinfectant solution, disconnect the
connecting tubes from the device. Otherwise, the disinfectant
solution may spray out of the tube, causing a leak from the
reprocessing basin.
Draining through the disinfectant collection hose
Do not push the disinfectant removal port when the rubber
cap is not attached. To do so may cause, the disinfectant
solution to flow out.
To prevent the device and areas near the equipment from
being damaged by leaking disinfectant solution, do not
remove the rubber cap from the disinfectant removal port
except when the disinfectant removal tube is connected.
Drainage disinfectant solution
Drainage volume Approximately 17.5 L (4.6 gallons)
Table 7.12
Check Required items
Disinfectant collection hose
Disinfectant removal tube
Containers (large) with 18 L (5 gallons) or larger capacity such as PVC
tanks (2)
Container (small) with 200 ml or larger capacity (wide-mouthed container such
as a beaker)
Clean cloth
Table 7.13