EasyManuals Logo
Home>Olympus>Laboratory Equipment>OER-Pro

Olympus OER-Pro User Manual

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
170
Chapter 6 Other Functions
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6.8 Rinsing
If cleaning fluid or disinfectant solution remains in the reprocessing basin or in
the endoscopes after an irregularity occurs or the process is stopped midway,
perform the rinsing process after the air purge process.
After disinfection, a water rinse is performed three times. Rinse conditions such
as number of rinsings performed and rinsing time cannot be changed. Worst
case time and water volume and water of each rinse cycle are shown below.
Time and water volume can vary depending on endoscope models, number of
endoscopes loaded and water supply conditions.
Be sure to perform the air purge first if cleaning fluid or
disinfectant solution is still in the reprocessing basin or
endoscopes. Otherwise, rinsing may be insufficient.
Before starting the rinsing, always confirm that there is no
foreign material on the ventilation openings on the gas filter
cases. Blocking the ventilation not only hinders deodorization
but may also cause the device to malfunction.
1. Close the lid.
2. Press the FUNC SEL button on the subcontrol panel to select “RINSE”.
Figure 6.21
Number of rinses 123
Time 100 s 110 s 130 s
Rinse water volume 12.4 L
3.3 gallons
14.9 L
3.9 gallons
14.9 L
3.9 gallons
Table 6.6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OER-Pro and is the answer not in the manual?

Olympus OER-Pro Specifications

General IconGeneral
BrandOlympus
ModelOER-Pro
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals