224
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
7. Turn the connectors to correct the orientations of the tubes as shown below.
Figure 7.55
8. Close the detergent/alcohol drawer.
7.10 Checking cassette cutters
This procedure is applied only when using concentrated
disinfectant prepared by cassette loading. When using
ready-to-use disinfectant prepared by top loading, the drawer
is locked.
Inspect the cassette cutters when replacing the disinfectant solution. For details,
see Section 7.12, “Replacing the disinfectant solution”.
7.11 Cleaning the disinfectant bottle drawer
This procedure is applied only when using concentrated
disinfectant prepared by cassette loading. When using
ready-to-use disinfectant prepared by top loading, the drawer
is locked.
It is necessary to clean the disinfectant bottle drawer if the disinfectant bottle
drawer is dirty or any disinfectant solution spills into the drawer. The disinfectant
solution bottle drawer can be opened only when replacing the concentrated
disinfectant solution. For details, see Section 7.12, “Replacing the disinfectant
solution”.