Chapter 3 Inspection Before Use
43
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
3.4 Inspecting the connecting tubes and leak test air
tube
Before using the equipment, always check that there is no irregularity regarding
the following points on the connecting tubes and leak test air tube.
All tubes should be free of cracks, breaks, fissures,
scratches, or stains.
There should be no cracks in the lock levers of connecting
tube connectors.
There should be no bends or breaks in the pin of
connecting tube connector.
The tube should not be easy to disconnect once
connected.
Figure 3.2
If a tube has any irregularity, do not use it and replace with a new one.
Do not use the connecting tubes or leak test air tubes if they
have any irregularity. Doing so could prevent effective
reprocessing or damage the endoscope.
Lock lever
Pin