EasyManuals Logo
Home>Olympus>Laboratory Equipment>OER-Pro

Olympus OER-Pro User Manual

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
Chapter 4 Reprocessing Operations
77
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If the lamp in the power switch does not turn ON
If the lamp in the power switch does not turn ON, check the equipment with the
following procedure. If the problem still persists after checking, set the power
switch to OFF and contact Olympus.
Before removing the fuse box, be sure to set the power
switch to OFF and unplug the power cord from the connector
on the equipment and the hospital-grade power outlet.
Otherwise, a fire or an electric shock may result.
To prevent an electric shock, do not check or inspect the
equipment with wet hands.
1. Ensure that the power cord is connected securely to the connector on the
equipment and to the hospital-grade power outlet.
2. Set the power switch to OFF and unplug the power cord from the
hospital-grade power outlet.
3. Unplug the power cord from the power cord receptacle on the equipment.
4. Push the tabs on the fuse box in the directions shown and take out the fuses
(see Figure 4.4).
Figure 4.4
(1)
The numbers in the figure indicate the fuse box
removal sequence.
(1)
(2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OER-Pro and is the answer not in the manual?

Olympus OER-Pro Specifications

General IconGeneral
BrandOlympus
ModelOER-Pro
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals