216
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
17. Close the lid.
18. Attach the gas filter case to the dome part of the lid.
19. Perform the operation described in Section 6.7, “Air purge” on page 169 to
remove water from inner piping of the equipment.
20. Step on the foot pedal to open the lid and take the sampling tube out of the
reprocessing basin, wipe off any water using a piece of clean cloth, and
store it in a clean place.
7.5 Replacing the air filter (MAJ-823)
Replace the air filter every month.
1. Open the front door of the device.
2. Remove the old air filter by pushing the sleeves on the two connectors
toward the device.
Figure 7.45
The air filter is not installed on the device when it is shipped
from the factory.
Check Required items
Air filter (MAJ-823)
Indelible pen or other writing implement
Table 7.9
FLOW
Sleeves