Appendix
291
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Specifications
Compatible
scopes
Olympus flexible endoscopes
(Refer to the “List of Compatible Endoscopes/Connecting Tubes
<OER-Pro>” for details)
(Note 1) The OER-Pro does not sterilize endoscopes.
Therefore, after cleaning/disinfecting an endoscope
that requires sterilization, always sterilize the
endoscope as instructed in its instruction manual.
Number of
reprocessed
scopes
Max. 2 (1 with certain models)
Cleaning method Exterior surfaces Ultrasonic cleaning, turbulent bath
Channel interiors Fluid flushing
Valves Ultrasonic cleaning, fluid flushing
Disinfection
method
Exterior surfaces Disinfectant solution immersion
Channel interiors Disinfectant solution flushing and filling
Valves Disinfectant solution immersion
Cleaning time
setting
3 – 10 minutes (Setting variable in
1 min. increments)
Disinfection time
setting
Concentrated
disinfectant solution
7 minutes
1
Ready-to-use
disinfectant solution
10 minutes
1
Disinfectant
solution heating
setting
20C (68F)
1
(If the temperature of disinfectant
solution is below 20C, it is heated to
20C (68F).)
Disinfectant
solution heating
method
Built-in heater in the reprocessing
basin.
1. Heating immediately before
disinfection process in a
reprocessing program
2. Heating before the start of a
reprocessing program
Water discharge
method
Forced draining using a pump (Floor
draining)
Disinfectant
solution discharge
method
1. Draining through disinfectant
collection hose
2. Draining through drain hose
Reprocessing
basin capacity
Approximately 14 L