192
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
7.2 Replacing the water filter (MAJ-824)
Replace the water filter at least once a month to prevent contamination of the
rinse water. The water filter should also be replaced whenever an error code
indicating water supply insufficiency [E01] is displayed. The water filter should be
replaced by following the flow shown below.
• Filter life varies depending on a number of factors including
incoming water quality and use volume.
• Using at least a prefilter (0.45 micron or less) can extend the
life of the water filter. If the prefilter is properly installed, the
water filter and the prefilter should be replaced at least once
every 6 months. For information on the Water Pre-filtration
system, contact Olympus.
Figure 7.8
Check Required items
Water filter (MAJ-824)
Water filter wrench
Water filter tubes ( 2)
Container with 2 L or larger capacity (wide-mouthed container such as a vat)
Indelible pen or other writing implement
Table 7.5
Replace the water filter
Drain water from the water filter housing
Disinfect the water supply piping
Drain air in the water filter housing