Chapter 3 Inspection Before Use
49
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If alcohol flushing is initiated without alcohol in the tank, the
error code [E93] will be displayed and the process will be
stopped.
When reprocessing is performed with newly-installed
equipment or with equipment after Section 7.16, “Preparing
the reprocessor for long-term storage” on page 259 is
performed, reprocessing may stop with the error code [E95]
displayed even though there is enough detergent in the tank.
See page 174 (“If error code [E93] is displayed while alcohol
is still present in the alcohol tank:”) to solve this problem.
Inspection of the amount of alcohol
Check the ALCOHOL indicator on the detergent/alcohol drawer of the equipment
to confirm that alcohol is present. If the amount of alcohol has decreased, add
alcohol as described in “Addition of alcohol” before the level reaches “MIN.”.
Figure 3.10
If alcohol flushing is executed when no alcohol is present, the
equipment will stop and display an error code [E93].

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OER-Pro and is the answer not in the manual?

Olympus OER-Pro Specifications

General IconGeneral
Power SupplyAC 100-240 V, 50/60 Hz
DisinfectionPeracetic acid-based disinfection
CompatibilityOlympus flexible endoscopes
Cleaning StagesDisinfection, Rinse
TypeAutomated Endoscope Reprocessor

Related product manuals