Chapter 4 Reprocessing Operations
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
To light the PRINT LAST and PRINT TODAY indicators
simultaneously, first press the FUNC SEL button so that the
PRINT TODAY indicator lights and then press the FUNC SEL
button again.
Only the results of successfully completed reprocessing
events are printed. The results of processes that were
aborted due to an error code are not printed out.
If the printout is not satisfactory, printing can be restarted.
Press the FUNC SEL button on the subcontrol panel so that
both the PRINT LAST and PRINT TODAY indicators light up
and then press the FUNC START button.
The same information can be printed several times until the
next reprocessing process is started.
For the printed information about “PRINT ALL”, refer to
“PRINT FORMAT (PRINT TODAY, PRINT ALL)” on
page 147.
If error code [E94] is displayed during printing
1.
If the paper runs out while printing, the main control panel displays error
code [E94]. Press the STOP/RESET button on the main control panel to
clear the error code.
2. Replace the printer paper roll.
3. After replacement, press the FUNC SEL button on the subcontrol panel so
that both the PRINT LAST and PRINT TODAY indicators light up.
4. Press the FUNC START button on the subcontrol panel. The printer starts
printing all of the stored reprocessing results again from the beginning.
5. If you want to print only the information after the point where printing was
stopped with error code [E94], do not clear the error code and simply
replace the printer paper roll.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OER-Pro and is the answer not in the manual?

Olympus OER-Pro Specifications

General IconGeneral
Power SupplyAC 100-240 V, 50/60 Hz
DisinfectionPeracetic acid-based disinfection
CompatibilityOlympus flexible endoscopes
Cleaning StagesDisinfection, Rinse
TypeAutomated Endoscope Reprocessor

Related product manuals