164
Chapter 6 Other Functions
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
When using Olympus-validated concentrated disinfectant
solution
When using Olympus-validated ready-to-use disinfectant
solution
• When using Olympus-validated concentrated disinfectant
solution, the “disinfection operation count” is set to 15 and
the “elapsed day count” is set to 3 at the factory.
• When using Olympus-validated ready-to-use disinfectant
solution, the “disinfection operation count” is set to 15 and
the “elapsed day count” is set to 14 at the factory.
Setting range
Disinfection operation
count
No setting or 1 – 35 times (Setting possible per time)
Elapsed day count No setting or 1 – 5 days (Setting possible per day)
Table 6.4
Setting range
Disinfection operation
count
No setting or 1 – 99 times (Setting possible per time)
Elapsed day count No setting or 1 – 14 days (Setting possible per day)
Table 6.5