182
Chapter 6 Other Functions
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
3. After pressing the ” button for 5 seconds, the “PRINT LAST” function
lamp of subcontrol panel starts to blink for indicating the current setting of
“AUTOMATIC PRINTING”. Make sure the indication of function lamp on the
subcontrol panel and determine the setting of “AUTOMATIC PRINTING” in
accordance with the description of Table 6.10. If you want to change the
setting, keep pressing the “ ” button and wait for the “PRINT TODAY”
function lamp of the subcontrol panel to start blinking. This indicates that
“AUTOMATIC PRINTING” has been deactivated. The setting will be
changed every 5 seconds by rotation.
The “Automatic printing function” is set to “Activated” at the
factory.
Panel display of sub-control panel
Setting of
AUTOMATIC PRINTING
Activated
Deactivated
Table 6.10
When “PRINT LAST” lamp blinks
When “PRINT TODAY” lamp blinks

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OER-Pro and is the answer not in the manual?

Olympus OER-Pro Specifications

General IconGeneral
Power SupplyAC 100-240 V, 50/60 Hz
DisinfectionPeracetic acid-based disinfection
CompatibilityOlympus flexible endoscopes
Cleaning StagesDisinfection, Rinse
TypeAutomated Endoscope Reprocessor