Chapter 7 Routine Maintenance
187
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
• When inspecting or cleaning this equipment, always wear
appropriate personal protective equipment, such as eyewear,
face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long
enough so that your skin is not exposed.
• If you find any problems or observe an irregularity, do not use
the equipment and contact Olympus. If the device is used
when an irregularity is observed, the device may malfunction.
Water leakage, electric shock, burns, and/or fire may also
result.
• Alcohol is flammable and should be handled with extra care.
• Before handling the alcohol, carefully read the precautions
for use, become fully acquainted with all safety materials,
and use the alcohol as instructed.
Check Monthly maintenance
7.1 Replacing the gas filter (MAJ-822)
7.2 Replacing the water filter (MAJ-824)
1
7.3 Disinfecting the water supply piping
7.5 Replacing the air filter (MAJ-823)
7.6 Cleaning the float switch
7.7 Cleaning the detergent/alcohol drawer
7.8 Cleaning the accessories and accessory holder
Table 7.1
1 Using at least a prefilter (0.45 micron or less) can extend the life of the
water filter. If the prefilter is properly installed, the water filter and the
prefilter should be replaced at least once every 6 months. For information
on the Water Pre-filtration system, contact Olympus.
Check Weekly maintenance
7.9 Cleaning the alcohol tank
Table 7.2