210
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
20. Press the FUNC START button on the subcontrol panel once again.
Figure 7.34
21. Water flows continuously from the filter tube. In about 15 seconds, the
buzzer beeps indicating the end of the process (the CODE/TIME display on
the main control panel shows [- -]).
Figure 7.35
22. Disconnect the filter tube by pushing its lock lever.
23. Close the front door.
24. Step on the foot pedal to open the lid.
When disconnecting the water supply piping disinfection
hose from the connector, cover the end of the hose with a
hand wearing a waterproof glove so that the water inside the
hose does not splash.