Chapter 7 Routine Maintenance
219
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
2. Turn the float switch cover in the direction shown to remove.
Figure 7.49
3. Clean the float switch cover in running water. Remove dirt using a soft brush
with detergent.
4. While applying clean water to the float switch, clean the stem using the
brush. Move the float up and down manually and thoroughly clean around
the stem. Rinse the float switch by pouring clean water over it.
Figure 7.50
Float switch cover

Table of Contents