Chapter 3 Inspection Before Use
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6. Place the alcohol tank in the detergent/alcohol drawer so that the ventilation
tube of the alcohol tank sits on the front of the tray.
Placing the alcohol tank so that the ventilation tube sits on
the deeper side of the tray could damage the alcohol tank.
7. Connect the tube that was originally connected to the cap.
Figure 3.15
8. Turn the connector to correct the tube orientation as shown below. Confirm
that the tube is not bent.
Figure 3.16
9. Close the detergent/alcohol drawer.
Alcohol tank
Ventilation tube
Connector