5-14
OPERATION
5-10 Selecting an Input Type (K3MA-J/K3MA-L)
This parameter allows you to select an input type of the product from the following
for each product model.
K3MA-J
K3MA-L
Parameter Set value
Description
in-t
(IN-T)
0-20 0-20 mA
4-20 4-20 mA
0-5 0-5 V
1-5 1-5 V
5 ±5 V
10 ±10 V
Parameter Set value
Description
in-t
(IN-T)
0
Pt100
-200-850°C -300-1500°F
1 -199.9-500.0°C -199.9-900.0°F
2 0.0-100.0°C 0.0-210.0°F
3
JPt100
-199.9-500.0°C -199.9-900.0°F
4 0.0-100.0°C 0.0-210.0°F
5
K
-200-1300°C -300-2300°F
6 -20.0-500.0°C 0.0-900.0°F
7
J
-100-850°C -100-1500°F
8 -20.0-400.0°C 0.0-750.0°F
9
T
-200-400°C -300-700°F
10 -199.9-400.0°C -199.9-700.0°F
11 E0-600°C0-1100°F
12 L -100-850°C -100-1500°F
13
U
-200-400°C -300-700°F
14 -199.9-400.0°C -199.9-700.0°F
15 N -200-1300°C -300-2300°F
16 R
0
-1700°C0-3000°F
17 S 0-1700°C0-3000°F
18 B 100-1800°C 300-3200°F
Initial setting level
Resistance thermometer
Thermocouple
CHAPTER 5 OPERATION