Roche Diagnostics May 2009
Service Manual · Version 9.0 A-7
cobas b 221 system 1 Safety information
Disinfectants
Disinfectants
Deproteinizer
Composition Aqueous NaOCl solution with active chlorine ( 2%)
Hazards identification Due to the basic and oxidizing character of the reagent ("Deproteinizer") local
irritations after contact with eyes, skin or mucous membranes cannot be excluded.
First aid measures
Other disinfectants
Use standard alcohol-based (70%) disinfectants to clean the surface of the
instrument. Read the product information first.
o
Service and repair work must be done only as specified in this manual. Unqualified service or
repair work may result in warranty claims not being granted.
o
Use only suitable tools and test instruments for service and repair work.
o
Do not allow fluids to enter the interior of the instrument, because this may damage the
electronics.
o
Use only slightly moistened cloths or cotton sticks to clean the instrument.
Use only liquid disinfectant such as protein remover (Roche deproteinizer) or an alcohol-based
(about 70%) surface disinfectant.
Do not spray disinfectant directly onto the instrument because this could cause malfunctions in the
electronics.
Do not use any type of bleaching agent. Exception: Roche Deproteinizer.
Do not attempt to clean/decontaminate any part of the instrument before shutting it down and
unplugging it from the power source.
Before plugging in the instrument again and switching it on always wait for 15 minutes to allow the
disinfectant to evaporate.
After inhalation:
breath fresh air, drink large amounts of water
After skin contact:
wash with generous amounts of water, remove contaminated clothing
After eye contact:
rinse eyes with generous amounts of water, contact an eye specialist
After drinking:
drink large amounts of water, avoid vomiting, contact a doctor
Never use standard disinfectant to clean the tubes and tubing paths under any circumstances!
Do not use any type of bleaching agent. Exception: Roche Deproteinizer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 221 system and is the answer not in the manual?

Roche cobas b 221 system Specifications

General IconGeneral
BrandRoche
Modelcobas b 221 system
CategoryMedical Equipment
LanguageEnglish

Related product manuals