EasyManuals Logo
Home>Roche>Medical Equipment>cobas b 221 system

Roche cobas b 221 system User Manual

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #227 background imageLoading...
Page #227 background image
Roche Diagnostics May 2009
Service Manual · Version 9.0 D-21
cobas b 221 system 7 Troubleshooting
System stops
O Replace corresponding sample sensor.
e
For instructions, see Replacing the SS1, SS2, SS3 and SS4 optical sensors on
page B-39.
O Replace sample sensor board.
e
For instructions, see Replacing the sample sensor board on page B-91.
Error 10049 "Error SS2"
The optical sample sensor at the end of the sample inlet path could not be calibrated.
o Call up "Wash" to calibrate the sample sensor.
o Check the corresponding sample sensor in the sensor test: press [Calibrate] to
perform dry calibration. Then press [Inject bypass] / [Inject sample
distributor] and inject water through the input unit. Press [Measure]. The
measured value with liquid shoud be significantly lower than dry.
o Visually inspect the corresponding sample sensor for liquid residue.
Fluid present:
O Test corresponding valves and if necessary replace them.
e
For instructions, see Replacing the valves from sample distributor cartridge and Hb
cartridge on page B-83.
O Check tubing system for blockages and replace the relevant components.
No fluid present:
O Replace corresponding sample sensor.
e
For instructions, see Replacing the SS1, SS2, SS3 and SS4 optical sensors on
page B-39.
O Replace sample sensor board.
e
For instructions, see Replacing the sample sensor board on page B-91.
Error 10050 "Error SS3"
The optical sample sensor in the Hb cartridge could not be calibrated.
o Call up "Wash" to calibrate the sample sensor.
o Check the corresponding sample sensor in the sensor test: press [Calibrate] to
perform dry calibration. Then press [Inject bypass] / [Inject sample
distributor] and inject water through the input unit. Press [Measure]. The
measured value with liquid shoud be significantly lower than dry.
o Visually inspect the corresponding sample sensor for liquid residue.
Fluid present:
O Test corresponding valves and if necessary replace them.
e
For instructions, see Replacing the valves from sample distributor cartridge and Hb
cartridge on page B-83.
O Check tubing system for blockages and replace the relevant components.
No fluid present:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 221 system and is the answer not in the manual?

Roche cobas b 221 system Specifications

General IconGeneral
BrandRoche
Modelcobas b 221 system
CategoryMedical Equipment
LanguageEnglish

Related product manuals