EasyManuals Logo
Home>Roche>Medical Equipment>cobas b 221 system

Roche cobas b 221 system User Manual

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #201 background imageLoading...
Page #201 background image
Roche Diagnostics May 2009
Service Manual · Version 9.0 C-9
cobas b 221 system 6 Maintenance
Decontamination
Cleaning of the measuring chambers BG, ISE and MSS
1
Remove the cobas b 221 system top cover.
2
Open the appropriate measuring chamber.
3
Open the electrode clamping lever in the measuring chamber and remove the
electrodes.
4
Clean the measuring chamber with a cloth moistened with disinfectant.
5
Plug the electrodes back into place and close the electrode clamping lever.
Close the measuring chamber and replace the top cover.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 221 system and is the answer not in the manual?

Roche cobas b 221 system Specifications

General IconGeneral
BrandRoche
Modelcobas b 221 system
CategoryMedical Equipment
LanguageEnglish

Related product manuals