App-14
IM DL35
O
-03EN
Appendix 2
Relationship between the Main Channel
Sample Rate, Sub Channel Data Update
Rate, and Acquisition Memory Writing Rate
16-CH Voltage Input Module
Rate at Which Data Is Written to the Acquisition Memory
Number of Sub Channels Used 1 2 3 or 4 5 to 8 9 to 16
Sub Channel Sample Rate
(Data Update Rate)
200 k 100 k 50 k 20 k 10 k
Main Channel
Sample Rate Setting (S/s)
100 M 500 k 500 k 500 k 500 k 500 k
50 M 500 k 500 k 500 k 500 k 500 k
20 M 500 k 500 k 500 k 500 k 500 k
10 M 500 k 500 k 500 k 500 k 500 k
5 M 500 k 500 k 500 k 500 k 200 k
2 M 500 k 500 k 500 k 200 k 100 k
1 M 500 k 500 k 200 k 100 k 50 k
500 k 500 k 100 k 100 k 50 k 20 k
200 k 200 k 100 k 50 k 20 k 10 k
100 k 100 k 50 k 20 k 10 k 5 k
50 k 50 k 10 k 10 k 5 k 2 k
20 k 20 k 10 k 5 k 2 k 1 k
10 k 10 k 5 k 2 k 1 k 500
5 k 5 k 1 k 1 k 500 200
2 k 2 k 1 k 500 200 100
1 k 1 k 500 200 100 50
500 500 100 100 50 20
200 200 100 50 20 10
100 100 50 20 10 5
50 50 10 10 5 2
20 20 10 5 2 1
10 10 5 2 1
5 5 1 1
: Some data will not be updated. If this occurs, previous data is displayed consecutively.
If you set the sample rate (the main channel sample rate setting) of this instrument to a rate that
exceeds the sub channel data update rate, all data within the sub channel data update interval will
be the same.

Table of Contents

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals