EasyManuals Logo
Home>ACUTRONIC>Medical Equipment>fabian

ACUTRONIC fabian User Manual

ACUTRONIC fabian
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
17 Technical specifications 96
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Essential performance
#
Essential Performance Function
Standard
Clause
Delivery of ventilation at the patient-connection port
within the alarm limits set by the operator;
3
or
the following alarm conditions shall occur as
appropriate:
Low battery as internal electrical power source
nears depletion;
Low and High Minute Volume;
Low and High PEEP;
Obstruction / Occlusion;
Low and High Air / Oxygen Inlet Pressure;
Air / Oxygen Gas Supply Failure;
or
a technical alarm condition shall occur.
ISO 80601-2-12:
2011
202.6.2.1.10
201.11.8.101.2
201.12.4.103
201.12.4.106
201.12.4.107
201.11.8.101.1
201.11.8.101.2
201.13.101/.102
201.101.1
Measurement accuracy of the oxygen respiratory gas
monitor (RGM) is maintained within the IFU
specification, ± (3% absolute);
or
Low and High Oxygen Level alarm conditions shall
occur, as appropriate;
or
a technical alarm condition shall occur.
ISO 80601-2-12:
2011
201.12.4.101
ISO 80601-2-55:
2018
201.12.1.101.1
208.6.1.2
201.11.8.101.1
Measurement accuracy of the SpO
2
monitor is
maintained within the IFU specification only within the
range 70 to 100%:
± 2% pediatric patients, no motion conditions;
± 3% pediatric patients, motion conditions, and all
neonatal patients;
or
Low and High SpO
2
Level alarm conditions shall occur,
as appropriate;
or
a technical alarm condition shall occur.
ISO 80601-2-61:
2017
201.12.1.101.1
208.6.1.2.101
201.11.8.101.1
201.12.4
201.13.101
3
To evaluate if essential performance is maintained, the following degradations shall not be allowed in any
normal use case:
component failures;
changes in programmable parameters or settings;
reset to default settings;
change of operating mode;
initiation of an unintended operation; and
Error of Delivered Volume (V
ti
) of individual breaths greater than 35% and error of the DELIVERED VOLUME
averaged over a one minute interval greater than 25%.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian
CategoryMedical Equipment
LanguageEnglish

Related product manuals