EasyManuals Logo
Home>ACUTRONIC>Medical Equipment>fabian

ACUTRONIC fabian User Manual

ACUTRONIC fabian
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
65 11 Ventilation modes
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
When ventilating a patient using a nasal interface in nCPAP or DuoPAP
modes, the clinician must make a compromise between providing additional
flow to compensate for a leak at the nose, or providing an alarm to warn of a
leak, particularly when the appropriate size prongs are small. For a patient
who is capable of spontaneous breathing without support, leak
compensation may be set to a high setting, in order to provide additional
flow to compensate for a leak at the interface between the prongs and the
nose. This will provide a more stable pressure support in the presence of
variable leak (for example, with patient movement). However; in this
scenario, the ventilator may fail to provide an alarm is the event of
decannulation.
For a patient who requires an alarm in the event of decannulation, leak
compensation should be set to a low setting. This will provide less stable
pressure support in the presence of variable leak at the nose, but a reliable
alarm for decannulation.
See Section 10: Ventilation parameters
If Inspiration time is set lower than 1 second, time maximum Pressure can
NOT be reached.
Depending on the Tube set, nCPAP Generator and Humidifier
O
2
Therapy mode
11.2.1 (High and Low flow oxygen therapy) HFNC
O
2
Therapy allows use of a continuous flow of blended gas, between 0 to 15 L/MIN in NEO and 0
to 30 L/min in PED mode. Nasal cannulas of various makes like F&P, Atom or similar can be used.
There are no alarm functions active in this mode, except for the set FiO
2
NOTE:
This mode can also be used to put the ventilator in Standby mode. By setting a Flow of 4
L/min, the humidifier dual servo temperature controls remain active, so no need to switch
it OFF in case of short-term Standby mode.
There are no patient alarms active in this mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian
CategoryMedical Equipment
LanguageEnglish

Related product manuals