85 15 Sterilization / cleaning / disinfection
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
15.1.2 Manual disinfection options
Option A: CaviWipes Surface Disinfectant Wipes (EPA REG. NO.: 46781-14)
1. Unfold the CaviWipe and wipe the surface of the enclosure completely.
NOTE
: The surface treated must remain visibly moist for three minutes.
Use more wipes if necessary, to keep the surface moist for three minutes.
2. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one
minute.
3. Allow to dry in the air.
Option B: 10% bleach (Clorox EPA REG. NO.: 5813-1)
1. Prepare a 10% bleach solution using purified water.
Use a clean lint-free cloth dampened with the bleach solution.
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE
: The treated surface must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one
minute.
4. Allow to dry in the air.
Option C: 5% H
2
O
2
, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H2O2 solution using purified water.
Use a clean lint-free cloth dampened with the H2O2 solution
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE
: The treated surface must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
Accessories
15.2.1 Flow sensor cable
The flow sensor cable must be prepared after each patient treatment.
WARNING:
The flow sensor cable must NOT be sterilized under any circumstances.