8 Battery operation 60
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
8 Battery operation
The device features an internal battery.
Battery Run Time in DUOPAP mode on full charge is approximately 3 hours.
These values may vary depending on the parameter settings. The device continuously calculates
the remaining Run Time and display it in the Information bar.
The Battery charge is continuously monitored.
WARNING
: A reduction in Battery life when changing the Operating mode can
cause an Automatic shutdown of ventilator!
Changing the Operating mode can greatly shorten the battery life.
• Monitor the remaining Battery life when changing the settings.
• NEVER leave the device and patient unattended during battery
operation.
• Establish mains power supply in good time!
In the event of Power Failure, the pneumatic system will automatically open
against atmospheric pressure so that pressure can’t build in the breathing
system, allowing spontaneous patient breathing.
When storing the ventilator for long term storage, please remove the Battery
pack.