18 Electromagnetic compatibility statement
NOTE:
fabian Therapy evolution is a MEDICAL APPLIANCE subject to specific precautionary
measures regarding EMC and must be installed and started up according to the notices
and the instructions in this manual.
WARNING:
Portable medical HF communication devices could impact fabian Therapy evolution.
WARNING:
NEVER use fabian Therapy evolution directly adjacent to, or stack with other apparatuses.
If unavoidable, be sure to monitor the equipment for proper operation with this set-up.
WARNING:
Using non-manufacturer recommended accessories, cables or converters with fabian
Therapy evolution could result in increased electromagnetic interference and reduce the
immunity of fabian Therapy evolution.
NOTE:
The key performance characteristics of the fabian Therapy evolution ventilator are:
• fabian Therapy evolution must operate within the defined specifications and the
medical purpose. Failure to meet these specifications will result in respiration being
stopped.
• For this reason, a second stand-alone ventilation unit must always be available,
(for example: Resuscitation Bag).
Device with SN from AT-04000
The "fabian Therapy evolution" device uses HF energy
solely for internal operation. Its HF emission is thus very
low and interference with adjacent apparatuses is unlikely.
The fabian Therapy evolution device is suitable for use in
all other areas except residential areas and areas directly
connected to a public mains supply which also supplies
buildings used for residential purposes.