61 8 Battery operation
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Power Failure
In the event of a Power Failure the device
will automatically switch to Battery Power
without interruption.
A visual and audible message is triggered
“Power failure“
This message can be confirmed
with the Alarm Silence button.
The device will continuously monitor the remaining battery run time.
Additional messages will be triggered at a remaining Run Time of:
• 15 minutes
• 30 minutes
• 60 minutes, only if the remaining run time is less than 60 minutes during a Power
Failure
If the remaining Run Time is less than 1
minute a warning will appear prompting to
immediately switch to a Primary Power
Source.
After 60 minutes on Battery power, the
battery is drained and fabian will power
OFF.
Depleted Battery!
– Connect to Mains –
– Charge battery for approx. 5 hours –