11 Ventilation modes 64
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
11 Ventilation modes
The ventilator is intended for the following ventilation modes:
NCPAP
NIV mask ventilation
DUOPAP
NIV mask ventilation
O
2
Therapy
High and Low Flow Oxygen Therapy HFNC
nCPAP/ DUOPAP
In nCPAP/DUOPAP mode the patient spontaneously breathes from a mask or nasal prongs.
A NIV Trigger option can be additionally implemented in order to provide those modes with
Breath detection (Apnea monitoring with alarms) and triggered supported breaths (or
synchronized transition from low to high CPAP levels) in DUOPAP.
Depending on the generator in use, fabian will then automatically select the proper NIV trigger
sensor:
Flow sensor (neonatal) with Infant Flow/INSPIRE and Infant Flow LP generators
Pressure sensor with Medijet (by Medin) generator
Monitoring of SpO
2
and PCO
2
is always recommended.
nCPAP: Supplies a positive airway pressure with automatic leak compensation.
The maximum flow compensation is selectable by menu.
DUOPAP: Same as nCPAP but with the option of positive pressure ventilation with
adjustable frequency and inspiratory period.
This mode requires a special nCPAP patient set with nCPAP generator.
Before using the nCPAP/DUOPAP mode the correct system must be specified
in the specifications menu.
The following systems are currently can be used:
Infant Flow
Infant Flow LP
MediJet